Series 651 Environmental Chambers
Series 651 Environmental Chambers enable test teams and researchers to perform a wide array of mechanical tests in conditions that closely replicate the temperatures that materials and components experience in their actual operating environments, providing confidence in the operating range of a material in the real world.
Applications
- High Temperature
- Low Temperature
Test Specimens
- Composites
- Metals
- Polymers
- Plastics
Test Standards
- ASTM D3039
- ASTM D3518
- ASTM D5023
- ASTM D5024
- ASTM D5026
- ASTM D5418
- ASTM D5992
- ASTM D6272
- ASTM D7028
- ASTM D7264
- ASTM E21
- ISO 14125
- ISO 6892-2
- ISO 14129
- ISO 527-4
- ISO 527-5
- ISO 6721-4
- ISO 6721-5
- ISO 6721-12
- EN 2562
- EN 2597
- EN 2746
- EN 6031
Key Product Features
Accurate & Repeatable
Designed for streamlined repeatable testing at any temperature
Easy Setup
Designed to streamline repeatable test setup.
Software Controlled
Precise temperature management to regulate temperature from the system PC.
Temperature Versatile
Can test as low as -150°C (-200°F) or as high as 540°C (1000°F).
Full Product Information
Every aspect of Series 651 Environmental Chambers was designed using decades of MTS experience developing test systems for numerous industries. These environmental chambers may include:
- Circulating fan helps to ensure small specimen temperature gradients
- Removable “U-plug” sections allowing chambers to be moved into place after the specimen is mounted and instrumented
- Doors that open to 180° or can be lifted away for convenient specimen access
- Multi-panel, tempered glass windows are sealed to reduce fogging and frosting
- Internal light to illuminate the test area
- Mobile carts to prevent interference between chambers and the load frame
Model 651.05F-01
Works in a variety of general materials testing applications up to 350°C (662°F) with floor-standing models of the MTS Acumen test system.
Model 651.05F-02
Is ideal for general materials testing applications up to 350°C (662°F) with tabletop models of the MTS Acumen test system.
Model 651.06E-03
Handles general materials testing up to 315°C (600°F) with floor-standing MTS Landmark test systems.
Model 651.10E-04
Offers support for general materials testing up to 540°C (1000°F) with floor-standing MTS Landmark test systems.
Model 651.06E-04
Is designed to accommodate large specimens or specimens that require large grips within the chamber. Works up to 315°C (600°F) with floor-standing MTS Landmark test systems.
ENVIRONMENTAL CHAMBERS SPECIFICATIONS |
|
|
|
|
|
MODEL NUMBER |
651.05F-01 |
651.05F-02 |
651.06E-03 |
651.06E-04 |
651.10E-04 |
Load frame |
Acumen 1 Acumen 3 Acumen 12 |
Landmark Bionix Acumen 3 Acumen 12 |
Landmark |
Landmark |
Landmark |
Maximum temperature*** |
350°C (662°F) |
350°C (662°F) |
315°C (600°F) |
315°C (600°F) |
540°C (1000°F) |
Chamber part number |
058-052-704 (US Plug)
058-052-703 (WW Plug)
|
058-052-702 (Acumen, US Plug)
058-052-701 (Acumen, WW Plug)
058-052-602 (w/table top legs, US Plug)
058-052-601 (w/table top legs, US Plug) |
056-097-206 (works with FlexTest SE Controller)
056-097-205 (works with FlexTest GT Controller)
056-097-207 (works with'494 FlexTest Controller) |
056-097-005 (works with FlexTest SE Controller)
056-097-004 (works with FlexTest GT Controller)
056-097-008 (works with 494 FlexTest Controller) |
056-097-105 (works with FlexTest SE Controller)
056-097-104 (works with FlexTest GT Controller)
056-097-106 (works with 494 FlexTest Controller)
|
Cart part number¶ |
100-294-319# 100-297-346 |
100-294-319# 100-297-346 |
100-160-646 100-213-156** |
100-160-646 100-213-156** |
100-160-646 100-213-156** |
Test space (nominal) |
216 mm W, 228 mm D, 228 mm H (8.5 in W, 9 in D, 9 in H) |
286 mm W, 305 mm D, 453 mm H (11.25 in W, 12 in D, 17.8 in H) |
356 mm W, 432 mm D, 559 mm H (14 in W, 17 in D, 22 in H) |
356 mm W, 432 mm D, 813 mm H (14 in W, 17 in D, 32 in H) |
356 mm W, 432 mm D, 660 mm H (14 in W, 17 in D, 26 in H) |
Removable “U-Plug” sections |
Not included |
Included |
Included |
Included |
Included |
Typical application |
Material testing, including fatigue and fracture, tension, compression, flex/bend, etc. |
||||
Air temperature performance*
Heating
LN2 cooling
Temperature gradient †, ‡
|
Ambient to +350°C (+660°F) in 45 minutes
Ambient to -150°C (-238°F) in 45 minutes
±3°C or less
|
Ambient to +350°C (+660°F) in 45 minutes
Ambient to -150°C (-238°F) in 45 minutes
±3°C or less
|
Ambient to +315°C (+600°F) in 30 minutes
Ambient to -129°C'(-200°F) in 26 minutes
±2°C (±5°F)
|
Ambient to +315°C (+600°F) in 30 minutes
Ambient to -129°C (-200°F) in 26 minutes
±2°C (±5°F)
|
Ambient to +540°C (+1000°F) in 70 minutes
Ambient to -129°C (-200°F) in 26 minutes
±2°C (±5°F) from -129° to 260°C (-200° to 500°F)
±5°C (±10°F) from -260° to 540°C ±1°C (±3°F)
|
Cooling requirements |
|
|
|
|
|
LN2 pressure |
0.15 MPa (22 psi nominal) 0.21 MPa (30 psi max) |
0.15 MPa (22 psi nominal) 0.21 MPa (30 psi max) |
0.15 MPa (22 psi nominal) 0.34 MPa (50 psi max) |
0.15 MPa (22 psi nominal) 0.34 MPa (50 psi max) |
0.15 MPa (22 psi nominal) 0.34 MPa (50 psi max) |
Recovery time to set point temperature Door open 1 minute Door open 2 minute |
5 minutes 6 minutes |
4 minutes 5 minutes |
8 minutes 10 minutes |
||
Heater circuit power |
208/230 V AC |
208/230 V AC |
208/230 V AC |
208/230 V AC |
208/230 V AC |
Requirements |
50/60 Hz, single-phase; 20 A circuit |
50/60 Hz, single-phase; 20 A circuit |
50/60 Hz, single-phase; |
50/60 Hz, single-phase; |
50/60 Hz, single-phase; |
Weight |
82 kg (180 lbs) |
108 kg (250 lbs) |
191 kg (4200 lbs) |
200 kg (440 lbs) |
195 kg (430 lbs) |
Internal light |
None |
Internal 120 V bulb (40 watt recommended) |
Internal 110 V bulb (220 V optional) (40 watt recommended) |
Internal 110 V bulb (40 watt recommended) |
Internal 110 V bulb (40 watt recommended) |
* Performance data derived with chamber empty and access holes blocked. Consult MTS Systems Corporation for temperature performance for testing specimens having high thermal mass. |
|||||
** For Landmark 370.50 only. |
|||||
***Typical maximum temperature for general materials testing applications. Many factors, including thermal mass of grips and fixtures, impact maximum operating temperature. |
|||||
¶ Consult MTS for cart information specific to your configuration. |
|
|
|
||
# Cart stand for table top load units. |
|
|
|
|
|
† After 30 minute stabilization time. |
|
|
|
|
|
‡ Due to large variety of possible test setups an air gradient is given. Specimen gradient is much less. |
|
|
Get The Most Out Of Your Investment
Our experts are here to help keep you up and running.
Related Products, Parts or Accessories
Acumen® Electrodynamic Test Systems
Landmark® Servohydraulic Test Systems
Bionix® Tabletop Test Systems
Criterion® Electromechanical Test Systems
Exceed® Electromechanical Test Systems
Looking for more products?
Go to Solution Finder